Based on Goldmann applanation tonometry and designed for use with the Takagi slit lamps. The AT-2 (Z-type) provides rapid, safe, and accurate intraocular pressure (IOP) measurement required for the diagnosis of such conditions as glaucoma, ocular hypertension, and retinal detachment.
Conforms to international standard ISO 8612:2009, which sets out requirements concerning accuracy and reliability of the measurements of applanations tonometers.
Finished in the same colour as the slit lamp to give a sense of uniformity when installed. Optional tonometer base TB2 is required for installation.
Measuring Prism - Smart integrated structure reduces risk of infection.
Measuring Prism S12-18 (included) features an integrated structure that reduces infection risk. Its environmentally conscious design is ideal for re-use and enables highly accurate and safe intraocular pressure measurement. Methacrylic resin, one of the hardest and most highly transparent optical-use plastic materials, is used to realise brightness and clarity of observation and to maximise product life.
Simply insert the disinfected measuring prism into the prism holder and operate the slit lamp joystick to make contact with and measure the cornea.
Attachable slit lamp models: | 2ZL, 2ZL BG, 4ZL |
Mounting tonometer base: | TB2 |
Measurement range: | 0-80 mmHg |
Minimum scale value: | 2 mmHg |
Measurement method: | Goldmann method (Goldmann Applanation Tonometry) |
Measurement deviation: | Nominal value ±1.5% or ±0.49 mN, whichever greater |
Reverse span (Hysteresis) | ≤0.29 mN |
Diameter of applanation field: | 3.06mm |
Depression area: | 7.354 mm2 |
Equation for conversion from mmHg to kPa: | 1 mmHg = 101325 / 760 Pa 133.3 Pa 100 Pa = 0.1 kPa |
Dimnsions: | 84.3 (W) × 207.6 (D) × 200.6 (H) mm |
Weight: | 0.67 kg |